MOSAIQ Oncology Information System

Company
Elekta, Inc.
Recall Initiated
January 15, 2025
Posted
February 12, 2025
Recall Number
Z-1162-2025
Quantity
64 units
Firm Location
Atlanta, GA

Reason for Recall

Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

Distribution

US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela

Lot / Code Info

Software versions /UDI: 3.1.2.0 (01)07340201500026 3.1.3.0 (01)07340201500026; 3.1.3.1 (01)07340201500026; 3.1.3.2 (01)07340201500026; 3.1.3.3 (01)07340201500026; 3.1.3.4 (01)07340201500026; 3.2.1.0 (01)07340201500071; 3.2.1.1 (01)07340201500071; 3.2.1.2 (01)07340201500071; 3.2.1.3 (01)07340201500071; 3.2.1.4 (01)07340201500071

Root Cause

Software design

Action Taken

On January 15, 2025 Elekta issued a recall notification to affected consignees via Email. Elekta asked consignees to take the following actions. 1. Customers included in the scope of the Field Safety Notice, should not calculate shifts from 2D verification images taken at a non-zero patient support angle (couch angle). 2. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 3. Advise the appropriate personnel, working with this product, on the content of this letter. 4. You are required to either acknowledge receipt of this notification via the Elekta Care" Community or complete this form and return it to Elekta immediately upon receipt, but no later than within 30 days.