Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

This recall is currently active, issued April 11, 2025. It was issued by Philips North America Llc.

Company
Philips North America Llc
Recall Initiated
March 10, 2025
Posted
April 11, 2025
Recall Number
Z-1553-2025
Quantity
83 units
Firm Location
Cambridge, MA
Official Source
View on FDA website ↗

Reason for Recall

Potential for collimator to fall as a result of incorrect installation.

Distribution

Worldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federation, Slovenia, South Africa, Spain, Thailand, United Kingdom, Vietnam.

Lot / Code Info

REF 712203; UDI: (01)00884838095205(21); Serial No. 220025, 210023, 220171, 230103, 230152.

Root Cause

Under Investigation by firm

Action Taken

Consignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an inspection of devices by field service engineers. Engineers may also replace the collimator, if necessary. Consignees with questions are to call Philips Customer Care Solutions Center at 1-800-722-9377 from Monday through Friday (8:00 AM to 8:00 PM EST).

More recalls by Philips North America Llc

View all recalls by this company →