3M Unitek Transbond Plus Self-Etching Primer REF 712-090 (100 unit box) 712-091 (20 unit box) Transbond Plus Self Etching Primer is an orthodontic all-in-one etchant and primer that is used in the bonding process of orthodontic appliances to teeth.
- Company
- 3M Unitek Corporation
- Recall Initiated
- November 26, 2024
- Posted
- January 2, 2025
- Recall Number
- Z-0797-2025
- Quantity
- 5,873 boxes (380,180 individual L-pops)
- Firm Location
- Monrovia, CA
Reason for Recall
Due to an increase of complaints for bracket bond failures and skin irritation or blistering.
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of : Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican, Republic, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Hong Kong, Japan, Korea, Latvia, Lithuania, Malaysia, Mexico, Netherland, New Zealand, Norway, Poland, Romania, Singapore, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine, and the United Kingdom.
Lot / Code Info
Product Number: 712-090 UDI-DI code: 00652221109170 Lot Numbers: 11093917 11106533 11120254 11153742 11155372 11173412 Product Number: 712-091 UDI-DI code: 00652221109187 Lot Numbers: 11165780 11165793
Root Cause
Process control
Action Taken
On 11/26/2024, the firm send an URGENT Product Recall letter via email to customers informing them that Solventum has received complaints of bracket bonding failures and skin irritation or blistering. Due to a manufacturing issue there is a potential for an increased occurrence of skin irritation/blistering and for reduced bracket bonding strength. Customer are instructed to: 1. Examine their inventory and destroy any affected product lots. Customers may dispose of the product per their facility s requirements. 2. Complete the enclosed Customer Product Acknowledgement Form and e-mail the completed form to 3MFieldSafetyAction@montage72.com. Upon receipt of the completed form, Solventum will issue replacement of product. For questions or further assistance - contact Montage customer service at 1-800-397-8118.