Multichem P, Part Code 08P90-10
This recall is currently active, issued October 12, 2023. It was issued by Techno-Path Manufacturing Ltd. Fort Henry Business Park Ballina Tipperary Ireland.
- Recall Initiated
- August 25, 2023
- Posted
- October 12, 2023
- Recall Number
- Z-0106-2024
- Quantity
- 2964 units
- Official Source
- View on FDA website ↗
Reason for Recall
Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.
Distribution
US Nationwide distribution in the state of Illinois.
Lot / Code Info
UDI/DI 05391523440769, Lot Numbers: 04530621P, 04561021P, 04610523P
Root Cause
Unknown/Undetermined by firm
Action Taken
LGC Group notified their sole US consignee (distribution partner) of the recall on 08/25/2023 via email. The notice explained the issue and the risk with the request to notify their end users. The end user was advised to take additional care when handling the glass vials due to the potential of breakages upon receipt, while thawing, or while opening the vials.