8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System

This recall is currently active, issued October 25, 2023. It was issued by B Braun Medical Inc.

Company
B Braun Medical Inc
Recall Initiated
September 21, 2023
Posted
October 25, 2023
Recall Number
Z-0123-2024
Quantity
9771
Firm Location
Allentown, PA
Official Source
View on FDA website ↗

Reason for Recall

Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.

Distribution

Domestic distribution nationwide. Foreign distribution to Canada.

Lot / Code Info

UDI-DI: 04046964660887 Serial Numbers: 868497 -892669 ***Updated 8/7/24*** SNs added: 430056, 430119, 430155, 430243, 551518, 685947, 703196, 705968, 726936, 737729, 753364, 774163, 790843, 814074

Root Cause

Component design/selection

Action Taken

On September 21, 2023, the firm began notifying customers of the issue via "URGENT MEDICAL DEVICE CORRECTION" letters. BBMI will schedule service to update impacted devices. Customers were provided with stickers to identify affected units while waiting for the service to be completed. In the meantime, customers were informed of measures that may aid in continued use of the pumps, such as only using non-impacted devices for the delivery of high-risk medications and placing the impacted units in care areas that do not administer high risk medications. Customers may contact BBMI by calling 1-800-627-7867 (1-800-627-PUMP) for technical support for potential alternate interim measures. ***Updated 8/7/24*** On August 7, 2024, firm identified additional units which were serviced using potentially impacted sensors. Serial numbers were added to the two models already in scope (Article Numbers 8713051U and 8713052U). Serial numbers from an additional model (Article No. 8713050U) also added and tracked under RES 95093.