RecallDepth

Easy Read App associated with First Response Pregnancy Test Sticks - Intended use of converting analog pregnancy test results into text pregnant or not pregnant for display on consumer devices (e.g., smart phone, iPad)

Company
Church & Dwight Inc
Recall Initiated
January 25, 2023
Posted
April 7, 2023
Recall Number
Z-1350-2023
Quantity
780000 downloads
Firm Location
Princeton, NJ

Reason for Recall

Church & Dwight First Response Pregnancy Kits was marketed with the Easy Read App without a new 510k. There is also a potential for Easy Read App to be unable to reads the test results (i.e., no results reported on the APP screen) or to misread the analog results shown on the pregnancy test stick and may lead to misinterpretation of false positive or false negative tests by the user.

Distribution

US Nationwide distribution.

Lot / Code Info

All lots with the Easy Read App

Root Cause

No Marketing Application

Action Taken

Church & Dwight Company Easy Read App requested the developer to remove the App from the Apple and Google Play app stores on 1/25/23 and third-party vendor disabled the App on consumers devices. Consumers who try to use the downloaded App receive a message that the App is no longer available.