CarePoint 60 mL Syringe Without Needle, REF: 04-7560
This recall has been completed (originally issued December 13, 2022).
- Company
- Allison Medical, Inc.
- Recall Initiated
- September 3, 2020
- Posted
- December 13, 2022
- Recall Number
- Z-0478-2023
- Quantity
- 2,004 Boxes
- Firm Location
- Littleton, CO
- Official Source
- View on FDA website ↗
Reason for Recall
Syringe without needle may have a green substance on the plunger. Substance found to be nontoxic but its effect on drug potency or interaction is unknown.
Distribution
US: IA, LA, MS, FL, PR; OUS: BARBADOS
Lot / Code Info
UDI-DI: 00786227756059, REF/Lot: 04-7560/ 80522
Root Cause
Nonconforming Material/Component
Action Taken
On 9/3/20, and later on 3/2/21, recall notices were emailed to customers who were informed to stop using and destroy or return affected devices to the recalling firm. Customers with additional questions could contact the recalling firm at 801-510-8733.