Humidifier bottle and syringe filter kit (6 count of each) for the BT37 Mark I Benchtop Incubator (AY102295). The Mark I model is no longer produced, but the legacy humidifier bottle component design can also be used with the BT37 Mark II Benchtop Incubator.
- Company
- Coopersurgical, Inc.
- Recall Initiated
- August 12, 2022
- Posted
- September 28, 2022
- Recall Number
- Z-1822-2022
- Quantity
- 5,182 kits
- Firm Location
- Trumbull, CT
Reason for Recall
There is a potential breach to the sterile barrier containing the Humidifier Bottles due to the packaging of the filter box located inside the bottle packaging. Damage to the sterile barrier may result in use of an unsterilized device, which may cause contamination and degradation or loss of embryo during incubation.
Distribution
Domestic distribution to the following states: AL AZ CA CT DE FL GA HI IL KS MA MI MN MO NJ NY OH OR SC TX VA VT WA WI WV Foreign distribution to Barbados, Canada, Chile, Micronesia Federate, and Panama.
Lot / Code Info
Lot Numbers 05593V291121 18-1192 1812-68 18-1268 G000575 G003811 G003893 G004517 G004874 G005109 G005476 G005723 E160415 Expiration date range: 27 Jun 2022 to 07 Dec 2024
Root Cause
Under Investigation by firm
Action Taken
On August 12, 2022, the firm notified customers via an Urgent Medical Device Safety Notice. Customers were instructed to quarantine affected product in their inventory. Once customers complete and return the Acknowledgement and Receipt Form attached to the recall letter, the firm will arrange for the affected product to be returned at no cost. Customers will receive credit for returned affected product. CooperSurgical may be reached at 203-601-5200 ext. 03300 or recall@coopersurgical.com.