Lunderquist -Ring Torque Wire Guide, Guidewire Reference Part Number/GPN THSF-35-125-THG G27031 THSF-35-145-THG G27033
This recall is currently active, issued September 28, 2022. It was issued by Cook Incorporated.
- Company
- Cook Incorporated
- Recall Initiated
- August 16, 2022
- Posted
- September 28, 2022
- Recall Number
- Z-1845-2022
- Quantity
- No distribution (Distribution Center only)
- Firm Location
- Bloomington, IN
- Official Source
- View on FDA website ↗
Reason for Recall
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Lot / Code Info
THSF-35-125-THG G27031 UDI-DI: 00827002270312 Lot/Expiration Date: 14804788 21-06-2027 THSF-35-145-THG G27033 UDI-DI: 00827002270336 Lots/Expiration dates: 14804789 21-06-2027 NS14804743 21-06-2027
Root Cause
Under Investigation by firm
Action Taken
On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235