Lunderquist -Ring Torque Wire Guide, Guidewire Reference Part Number/GPN THSF-35-125-THG G27031 THSF-35-145-THG G27033

This recall is currently active, issued September 28, 2022. It was issued by Cook Incorporated.

Company
Cook Incorporated
Recall Initiated
August 16, 2022
Posted
September 28, 2022
Recall Number
Z-1845-2022
Quantity
No distribution (Distribution Center only)
Firm Location
Bloomington, IN
Official Source
View on FDA website ↗

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Lot / Code Info

THSF-35-125-THG G27031 UDI-DI: 00827002270312 Lot/Expiration Date: 14804788 21-06-2027 THSF-35-145-THG G27033 UDI-DI: 00827002270336 Lots/Expiration dates: 14804789 21-06-2027 NS14804743 21-06-2027

Root Cause

Under Investigation by firm

Action Taken

On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235