RecallDepth

Rosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623 THSCF-35-80-1.5-ROSEN G01774 THSCF-35-180-1.5-ROSEN G01264 THSCF-35-145-1.5-ROSEN G01261

This recall is currently active, issued September 28, 2022. It was issued by Cook Incorporated.

Company
Cook Incorporated
Recall Initiated
August 16, 2022
Posted
September 28, 2022
Recall Number
Z-1842-2022
Quantity
5041 US; 298 OUS
Firm Location
Bloomington, IN
Official Source
View on FDA website ↗

Reason for Recall

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Distribution

Domestic distribution nationwide. Foreign distribution worldwide.

Lot / Code Info

THSCF-35-220-1.5-ROSEN G01623 UDI-DI: 00827002016231 Lots/Expiration Date: 14818344 29-06-2027 14820965 30-06-2027 14849857 20-07-2027 THSCF-35-80-1.5-ROSEN G01774 UDI-DI: 00827002017740 Lots/Expiration Date: 14820960 30-06-2027 14832877 08-07-2027 14847945 19-07-2027 THSCF-35-180-1.5-ROSEN G01264 UDI-DI: 00827002012646 Lots/Expiration Date: 14820961 30-06-2027 14843133 15-07-2027 14847946 19-07-2027 THSCF-35-145-1.5-ROSEN G01261 UDI-DI: 00827002012615 Lots/Expiration Date: 14820962 30-06-2027 14820963 30-06-2027 14820964 30-06-2027 14828587 06-07-2027 14835364 11-07-2027 14835365 11-07-2027 14849859 20-07-2027

Root Cause

Under Investigation by firm

Action Taken

On August 16, 2022, Cook Medical notified affected customers of the recall. Customers were instructed to immediately cease all further distribution and use of the affected products. Affected product should be quarantined and returned to Cook Medical to receive product credit. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235

More recalls by Cook Incorporated

View all recalls by this company →