INTERA 3000 Hepatic Artery Infusion Pump, 30 mL, Sterile, Rx Only. implantable infusion pump.

This recall has been terminated (originally issued August 22, 2022).

Company
Intera Oncology, Inc.
Recall Initiated
July 10, 2022
Posted
August 22, 2022
Terminated
September 12, 2023
Recall Number
Z-1522-2022
Quantity
50 units
Firm Location
Wellesley, MA
Official Source
View on FDA website ↗

Reason for Recall

Higher than expected flow rate.

Distribution

US distribution to states of: AL, CA, IL, KY, MN, MO, NC, NJ, NY, OH, OR, PA, & TX.

Lot / Code Info

UDI-DI: 00850014110147; Catalogue No. AP03000H; Serial No. 16145, 16151, 16154, 16171, 16177, 16179, 16196, 16201, 16204, 16211, 16212, 16219, 16220, 16235, 16244, 16247, 16260, 16270, 16273, 16298, 16304, 16309, 16313, 16319, 16338, 16346, 16354, 16376, 16379, 16389, 16393, 16394, 16395,16402, 16403, 16405, 16408, 16409, 16411, 16425, 16426, 16428, 16433, 16434, 16437, 16450, 16459, 16461, 16464, 16485.

Root Cause

Process design

Action Taken

Customers were notified by a combination of letter, call, and/or email of this recall on July 10, 2022. Customers were informed via notification that Intera 3000 Hepatic Artery Infusion Pumps are being recalled due to complaints reporting higher-than-expected flow rates. Customers are asked to secure affected pumps in inventory to ensure they are not implanted and to contact Intera Oncology at 1-800-660-2660 to arrange return of product. Intera requests physicians continue to monitor the pump flow rate per standard refill procedure, and continue to monitor for liver toxicity to adjust dosing as per standard protocols.