RecallDepth

GreenLight HPSEA Laser Fibers, UPN 0010-2092

Company
Boston Scientific Corporation
Recall Initiated
May 31, 2022
Recall Number
Z-1392-2022
Quantity
57
Firm Location
Marlborough, MA

Reason for Recall

Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.

Distribution

Domestic distribution to AR CA IL ME MI NC TX Foreign distribution to Australia, Canada, and France.

Lot / Code Info

UDI-DI (GTIN): 00878953005522 Lot: 27770466

Root Cause

Software Manufacturing/Software Deployment

Action Taken

On May 31, 2022, the firm distributed Urgent Medical Device Product Removal letters to affected customers. Customers were instructed to immediately segregate and return the product to Boston Scientific.

More recalls by Boston Scientific Corporation