Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031

Company
Philips North America Llc
Recall Initiated
November 24, 2021
Recall Number
Z-0725-2022
Quantity
1 unit
Firm Location
Cambridge, MA

Reason for Recall

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Distribution

US Nationwide Foreign: Argentina Austria Bangladesh Brazil Chile China Denmark Dominican Republic France Germany Greece India Indonesia Iran Israel Italy Japan Korea, Republic of Malaysia Mexico Netherlands Philippines Russian Federation Saudi Arabia South Africa Spain Sri Lanka Switzerland Taiwan Thailand Turkey United Kingdom Virgin Islands (U.S.)

Lot / Code Info

S/N: OUS 109

Root Cause

Process control

Action Taken

Philips Issued Medical Device Correction letter dated 11/24/21 to U.S.A, Customers via certified mailing from the United States Postal Service (USPS). Letters outside of the U.S.A. will be distributed through the Philips Markets Organizations. Letter states reason for recall, health risk and action to take: Philips will inspect the affected systems for leaks and repair as required. In addition, Philips will install an extended drip tray to prevent damage from any future leaks. This service will be provided free of charge for affected systems. A Philips service representative will contact customers with affected devices to arrange for the service. If you need any further information or support concerning this issue, please contact your local Philips representative: 1-800-722-9377 (reference FCC 72200484).