VITROS 350 Chemistry System, Product Code 6802153, (UDI # 10758750002054) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.

This recall has been terminated (originally issued April 8, 2019).

Company
Ortho-Clinical Diagnostics
Recall Initiated
April 8, 2019
Terminated
November 12, 2020
Recall Number
Z-1936-2019
Quantity
3889 units
Firm Location
Rochester, NY
Official Source
View on FDA website ↗

Reason for Recall

Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.

Distribution

US Nationwide Distribution

Lot / Code Info

All lots

Root Cause

Device Design

Action Taken

On 08 April 2019, a customer letter (Ref. CL2019-090) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) to either all customers who were shipped affected VITROS 250 or 350 Systems. This communication provided the following required actions: - As per the VITROS CK Slide Instructions for Use, perform Quality Control (QC) testing daily to help ensure the slides are performing within acceptable limits. - If QC results are outside of the established ranges, discard the cartridge and repeat QC testing using a fresh cartridge of VITROS CK Slides. - If multiple cartridges are loaded on the VITROS 250/350 System, daily QC testing must be performed for each cartridge prior to use. - Consider loading one cartridge of VITROS CK Slides on board your system at a time. - Complete and return the request form via fax to 1-888-557-3759. Foreign affiliates were informed by email on 08 April 2019 of the issue and instructed to notify affected consignees of the issue and appropriate actions to take.

More recalls by Ortho-Clinical Diagnostics

View all recalls by this company →