RecallDepth

Prometra Programmer (Grand Prime) Software version 2.00.29, Cat. No. 13828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.

Company
Flowonix Medical Inc
Recall Initiated
March 21, 2019
Terminated
June 5, 2020
Recall Number
Z-1713-2019
Quantity
40
Firm Location
Mansfield, MA

Reason for Recall

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.

Distribution

US nationwide distribution.

Lot / Code Info

Software v.2.00.29 Serial numbers: D18L0003 D18L0022 D18L0013 D18L0035 D18L0004 D18L0025 D18L0014 D18L0036 D18L0006 D18L0026 D18L0015 D18L0037 D18L0010 D18L0027 D18L0016 D18L0038 D18L0009 D18L0028 D18L0017 D18L0034 D18L0008 D18L0029 D18L0018 D18L0039 D18L0007 D18L0030 D18L0019 D18L0040 D18L0011 D18L0031 D18L0021 D18L0041 D18L0005 D18L0032 D18L0023 D18L0042 D18L0012 D18L0033 D18L0024 D18L0020

Root Cause

Software change control

Action Taken

On March 21, 2019, the firm issued Urgent Recall Notices to customers to inform them of the product issue. In May 2019, the firm sent updated communications to customers: Healthcare providers who have Clinician Programmers with SW 2.00.29 were sent a follow-up recall letter. This letter included the information that the pump stop following "Error Code 115" will only occur after the Bridge or Demand Bolus is complete, which could take up to seventy-two (72 hours). In addition, in the event of any cause of an "Error Code 115", the pump sounds a one-beep alarm and then shuts down. This behavior differs from the Instructions for Use, which state that the pump will sound three long (half second) beeps every 30 minutes for all critical errors. As it is unlikely that the patient will notice the single beep, there may be a significant delay in the patient becoming aware that the pump has stopped delivering therapy which may in turn delay resumption of drug therapy. Healthcare providers who have Clinician Programmers with any other version of the software and the implant pumps received a Corrective Action Notice. This notice included the information from the preceding paragraph, noting the discrepant alarm behavior for "Error Code 115" could be missed by the patient. Customers with the new Prometra Clinician Programmer software (v. 2.00.29) were asked to do the following: 1. STOP using the Periodic Flow or Multiple Rates flow mode functions in combination with a Bridge or Demand Bolus using the Clinician Programmer with Software Version 2.00.29. The error only occurs when these functions are used together. a. In the event that a Bridge or Demand Bolus is needed, convert the patient to Constant Flow Mode, then program the Bridge or Demand Bolus as needed (instructions for use PL-11803-04, page 30). Once the bolus is complete, revisit the patient or have him/her return to your clinic for programming back into their original flow mode. 2. IDENTIFY

More recalls by Flowonix Medical Inc

View all recalls by this company →