AURIGA XL 4007 GENERAL - ZERO COST SYSTE UPN: FS4007GZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties including: Urology, Urinary Lithotripsy, Gastroenterology, Arthroscopy, Discectomy, Pulmonary, Gynaecology, ENT, Dermatology, Plastic Surgery and General Surgery.

This recall has been terminated (originally issued May 22, 2019).

Company
Boston Scientific Corporation
Recall Initiated
April 8, 2019
Posted
May 22, 2019
Terminated
May 10, 2023
Recall Number
Z-1394-2019
Quantity
6 (OUS)
Firm Location
Marlborough, MA
Official Source
View on FDA website ↗

Reason for Recall

Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.

Distribution

Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam

Lot / Code Info

Serial Number:  21303005, 21303006, 21303043, 21303089, 21303066, 21303090

Root Cause

Component design/selection

Action Taken

Boston Scientific issued to the user level the correction/removal via a customer notification letter via FedEx or hand delivery by a Boston Scientific agent on 4/9/19. Letter provides recommendations to the user until a field repair can be completed or a replacement console can be provided. A Boston Scientific representative will be in contact with you directly to schedule a field repair or replacement at your facility. The console can continue to be used during this time following the recommendations within the letter.