RUSCH LASERTUBE (Rubber), Order No. : 102004-000080, REF 102004, SIZE I.D. mm 8,0
- Company
- Teleflex Medical
- Recall Initiated
- April 1, 2019
- Terminated
- June 3, 2025
- Recall Number
- Z-1686-2019
- Quantity
- 2325 units
- Firm Location
- Morrisville, NC
Reason for Recall
The firm received reports indicating that the laser guard foil partially separated and/or slightly detached at the edges.
Distribution
Nationwide
Lot / Code Info
Lot Numbers: 14151, 14161, 14201, 14281, 14301, 14321, 14431, 14491, 14511, 15041, 15371, 15441, 16071, 16131, 16191, 16221, 16241, 16271, 16311, 16371, 16391, 16441, 16451, 16481, 16491, 17091, 17121, 17181, 17211, 17241, 17281, 17351, 17381, 17451, 17471, 17491, 18181, 18301, 18411
Root Cause
Process control
Action Taken
The firm initiated the recall by letter on 04/01/2019. The letter stated reason for the recall and requested the return of the recalled units. Distributors were directed to collect the recalled units from their customers, consolidate them, and complete and return the acknowledgement form. The recalled units will be returned.