The Alinity ci-series System Control Modules which are configured with Alinity c Processing Modules. The Alinity ci-series Control Module is labeled in part,"* * *Alinity ci-series* * *SYSTEM CONTROL MODULE Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. The Alinity c Processing Module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c Processing Module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c Processing Module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).
This recall has been terminated (originally issued May 21, 2019).
- Recall Initiated
- May 21, 2019
- Terminated
- May 20, 2025
- Recall Number
- Z-2201-2019
- Quantity
- 943 systems
- Official Source
- View on FDA website ↗
Reason for Recall
All versions of the Alinity ci-series software may not detect an issue on the Alinity c Integrated Chip Technology (ICT) assays, sodium (Na+), potassium (K+), and chloride (Cl-).
Distribution
Worldwide Distribution - US Nationwide AL, AR, CA,, FL, GA, ID, IL, LA, MA, ME, MI, MN, MO, NJ, NY, OK, SC, TN, TX, UT, VA, WA, WI, and PUERTO RICO. ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM,BOTSWANA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLUMBIA, CROTIA, CZECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDOESIA, IRELAND, ISRAEL, JORDAN, KENYA, KUWAT, LATVIA, LEBONON, MEXICO, NAMIBA, NETHERLANDS, NORWAY, PACKISTAN, PHILLINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDON, and VIETNAM.
Lot / Code Info
List Number: 03R70-01 ALL LOTS
Root Cause
Under Investigation by firm
Action Taken
On May 21, 2019, the firm sent Product Correction Urgent - Immediate Action Required letters to all of their consignees which advised them of the problem and provided them with instructions for preforming additional testing. US customers were sent this letter and an attached Customer Reply Immediate Action Required form vis FedEx priority overnight express. Abbott service personnel will follow-up to install new software.