Regard custom surgical pack labeled as "LAPAROSCOPIC - OSU" including recalled Stryker part numbers 0620 040 660 Pneumosure High Flow II Tubeset, 0620 040 690 Pneumosure Heated Tubeset with RTP and 0620 040 680 Pneumosure High Flow Tubeset with RTP.

Recall Initiated
March 21, 2019
Recall Number
Z-2221-2019
Quantity
312 packs
Firm Location
Springfield, MO

Reason for Recall

Resource Optimization & Innovation (ROi) has received a recall notice from Stryker regarding tube sets potentially detaching from the cassette, which can lead to leaking gas or a hissing sound.

Distribution

Distributed to LA and MO

Lot / Code Info

Item Number (Lot Number, Expiration Date): 880341004 (043444D, 43566);  880341005 (047474E, 43706); 880341005 (049043E, 43750);

Root Cause

Nonconforming Material/Component

Action Taken

ROi, Consolidated Service Center ROi, notified customers on about 03/21/2019. Each email included a copy of the ROi "MEDICAL DEVCIE RECALL NOTIFICATION" dated March 21, 2019 as well as the Stryker recall notification dated March 7, 2019. ROi notification letter instructed customers to review inventory to determine if the affected ROi Custom Pack item and lot numbers are in inventory, quarantine all affected packs on hand, and notify ROi of the number of each affected lot on hand by completing and returning "Attachment A" via email to lacy.stewart@roiscs.com. Supplemental labels for affected packs will be issued, which will instruct the end user to remove the affected Stryker component(s) and to use single sterile replacement product form inventory. If there is not any affected inventory on hand, customers are requested to still return Attachment A indicating "0" affected inventory at their facility.