RecallDepth

Brilliance iCT, 728306 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories

This recall has been terminated (originally issued July 3, 2018).

Company
Philips Medical Systems (Cleveland) Inc
Recall Initiated
July 3, 2018
Terminated
April 17, 2020
Recall Number
Z-1144-2019
Quantity
2
Firm Location
Cleveland, OH
Official Source
View on FDA website ↗

Reason for Recall

Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.

Distribution

Distributed only to China and UK.

Lot / Code Info

85028   100094

Root Cause

Device Design

Action Taken

On June 29. 2018, the firm sent Customer Information Letters to affected customers. The letter informed customers that Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. Customers were advised to review the user documentation provided with your system related to Quality Assurance checks. If artifacts are detected during review of the images produced by these checks, and are not removed with the recommended steps identified within your user documentation, please contact your service representative before resuming clinical use.

More recalls by Philips Medical Systems (Cleveland) Inc

View all recalls by this company →