Instrument Tray, Basic T2 Femur Long Catalog #: 1806-9900 Product Usage: Knee Arthroplasty
- Recall Initiated
- January 21, 2019
- Terminated
- April 2, 2020
- Recall Number
- Z-0980-2019
- Quantity
- 17 units
Reason for Recall
The faceplates may incorrectly identify the T2 Basic Long Instrument Set as T2 Basic Short Instrument Set.
Distribution
In the countries of Canada and Korea
Lot / Code Info
Lot Number(s): K026ACE (lot# packaging) KU97181 (Vendor code engraved on device)
Root Cause
Employee error
Action Taken
Zimmer Biomet issued Urgent Field Safety Notice PFA 1982723 customer letter dated: January 21, 2019. Letter identifies affected product, problem and action to be taken: Inform individuals within your organization who need to be aware of this action. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. Response is required, even you may not have any physical inventory on site anymore. 3. Quarantine and discontinue use of the devices.