RecallDepth

Beijing Syntech Laser APOLLO V+ Medical Platform surgical lasers

Company
Beijing Syntech Laser Co., Ltd International Industry Base West Area
Recall Initiated
July 3, 2017
Recall Number
Z-1492-2018
Quantity
11
Firm Location
Changping Beijing China

Reason for Recall

The noncompliance is that the certification label was not affixed on the product, with statement "Complies with 21 CFR 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007".

Distribution

US Distribution

Lot / Code Info

Model Apollo V + Medical Platform

Root Cause

Nonconforming Material/Component

Action Taken

Beijing Syntech Laser measures to be taken: 1. Beijing Syntech Laser Co. Ltd. has sent notifications to all relevant dealers, distributors, and purchasers, 2. Beijing Syntech Laser Co. Ltd. will send certification labels and instructions to apply the label to the product to all relevant dealers, distributors, and purchasers, 3. The corrections will be conducted at no cost to the purchaser, and 4.The corrective action will be completed by August 1st, 2018. CDRH approves the CAP subject to the following condition: Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22, if it hasnt been already. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. For further questions please call (360) 325-7028

More recalls by Beijing Syntech Laser Co., Ltd International Industry Base West Area

View all recalls by this company →