LeMaitre Vascular Reddick Cholangiogram Catheter (eIFU) REF# e2400-50

Company
Lemaitre Vascular, Inc.
Recall Initiated
February 7, 2017
Terminated
September 20, 2019
Recall Number
Z-1442-2017
Quantity
5601
Firm Location
Burlington, MA

Reason for Recall

The Contains Latex symbol was not printed onto the box label or the sterile (inner) product pouch

Distribution

Worldwide distribution. US Nationwide, AUSTRIA, BELGIUM, FRANCE, GERMANY, ITALY, LUXEMBOURG, Morocco, Saudi Arabia, Singapore, SPAIN, SWITZERLAND, UAE, and UK

Lot / Code Info

Lot Number/Exp. Date: RED1743 2018-09 RED1744 2018-09 RED1749 2018-10 RED1750 2018-12 RED1751 2018-12 RED1755 2018-12 RED1756 2018-12 RED1757 2019-01 RED1759 2019-02 RED1760 2019-02 RED1761 2019-03 RED1764 2019-03 RED1766 2019-04 RED1767 2019-04 RED1768 2019-05 RED1769 2019-05 RED1771 2019-06 RED1773 2019-07 RED1774 2019-07 RED1775 2019-07

Root Cause

Labeling design

Action Taken

LeMaitre Vascular, Inc. issued a Dear Doctor letter, on 2/7/17, to distributors and hospitals via FedEx overnight. The letter contains a form that is requested to be returned to LeMaitre Vascular, Inc. as a record of notification and reconciliation. Any unused product requested to be returned and replaced. Questions concerning this notice contact 781-221-2266 ext. 183.