RecallDepth

Genetic Screening Processor (GSP); PerkinElmer Part Number= 2021-0010 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.

This recall has been terminated (originally issued January 27, 2017).

Company
Perkinelmer Life And Analytical Sciences, Wallac, Oy Mustionkatu 6
Recall Initiated
January 5, 2017
Posted
January 27, 2017
Terminated
August 24, 2017
Recall Number
Z-1107-2017
Quantity
24 instruments and 3 spare disk remover modules
Firm Location
Turku Finland
Official Source
View on FDA website ↗

Reason for Recall

The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.

Distribution

MA, SC, TX, AL OUS: China, Mexico, Peru, Morocco, Italy, Finland, Ecuador, Brazil

Lot / Code Info

Serial Numbers: 20210239 20210243 20210244 20210245 20210246 20210247 20210248 20210249 20210250 20210251 20210252 20210253 20210254 20210255 20210256 20210257 20210258 20210259 20210260 20210261 20210262 20210263 20210264 20210265

Root Cause

Component change control

Action Taken

The recall was initiated on January 5, 2017 via email sent through the PerkinElmer Technical Support group located in Akron Ohio. Actions to be taken by user: 1. As instructed in the analyte specific kit inserts, the use of Kit Controls is required to assure the day-to-day validity of results. Patient results should only be reported if control results for the assay meet the laboratorys established criteria for acceptability. 2. If a movement error occurs customers will be directed to contact their local PerkinElmer field service representative. The movement error may indicate deterioration of the axle and therefore the use of the instrument should be avoided until field service personnel have inspected the status of the ball bushing axles or replaced affected ball bushing axles. Actions to be taken by the PerkinElmer: As a corrective action, PerkinElmer will replace the affected ball bushing axles in all the GSP instruments identified in this FSCA/Recall. The replacement may take place either as part of the initial installation activity before the instrument is used, scheduled periodic maintenance, or during a dedicated service visit. This field correction is considered as mandatory by the manufacturer. Customers will be contacted by PerkinElmer to arrange the field service visit.

More recalls by Perkinelmer Life And Analytical Sciences, Wallac, Oy Mustionkatu 6

View all recalls by this company →