Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068

Company
Siemens Healthcare Diagnostics, Inc.
Recall Initiated
August 19, 2020
Terminated
August 25, 2023
Recall Number
Z-0052-2021
Quantity
533 US and 1822 OUS
Firm Location
Newark, DE

Reason for Recall

Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed

Distribution

US Nationwide distribution.

Lot / Code Info

Lot # EB0255 UDI: 00842768037726EB025520091110471068840

Root Cause

Under Investigation by firm

Action Taken

Siemens issued Urgent Medical Device Correction (UMDC # VC-20-03.B.US) and Urgent Field Safety Notice (UFSN # VC-20-03.B.OUS) issued on August 19, 2020 via FedX to US Dimension customers who received the Dimension TNIH reagent lot to notify them of the issue, the risk to health, and the following actions to be taken by the customer: Review the letter with their Medical Director"If they are using TNIH lot EB0255 DO NOT switch to another TNIH lot during serial measurements for a single patient sample. Discontinue use and discard the TNIH reagent lot EB0255 upon receipt of replacement product. Request your no-charge replacement for lot EB0255. Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. Questions, contact your Siemens Remote Services Center or your local Siemens Technical Support representative.