VITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.

Company
Ortho-Clinical Diagnostics
Recall Initiated
July 17, 2020
Terminated
July 16, 2021
Recall Number
Z-2771-2020
Quantity
2349 sales units
Firm Location
Rochester, NY

Reason for Recall

The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.

Distribution

Worldwide distribution - US Nationwide distribution.

Lot / Code Info

Lots 0010 - 0013

Root Cause

Process control

Action Taken

On July 17, 2020, the firm issued Urgent Product Correction Notification letters to affected customers. Customers were informed of the potential for erroneous patient sample results, calibration errors, and QC failures when using the impacted product. Customers were asked to do the following; " Visually inspect affected VITROS Anti-SARS-CoV-2 Total (Lots 0010-0037) and IgG (Lots 0010-0013) Reagent Packs to ensure correct reagent bottle configuration " Set aside any pack in which the reagent bottles are positioned incorrectly, so the lot number and pack id can be provided to Ortho. Complete and submit the enclosed Request for Credit Form, including the pack information, to receive credit for an affected pack. " Complete the Confirmation of Receipt form no later than July 28, 2020. " Post this notification by each system that processes an affected lot. " Please forward this notification if the product was distributed outside of your facility. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.

More recalls by Ortho-Clinical Diagnostics

View all recalls by this company →