VACUETTE Tube 2ml 9NC Coagulation Sodium Citrate 3.2%. Item no. 454322. Lot no. B1908355.

Company
Greiner Bio-One North America, Inc.
Recall Initiated
March 12, 2020
Terminated
August 31, 2022
Recall Number
Z-1714-2020
Quantity
1,000,800 units.
Firm Location
Monroe, NC

Reason for Recall

Concerned lot of VACETTE 2mL 9NC Coagulation Tubes were shown to have inconsistent amounts of the 2.3% sodium citrate.

Distribution

US nationwide distribution in the states of: NJ, IN, TN, MI, NC, WA, KS, CA, LA, IL, MD, NY, OH, TX, GA, KY, PA, CO, MA, PR, FL, AR, OK and NE.

Lot / Code Info

Item no. 454322. Lot no. B1908355. DI no. (01)29120017574488 (17) 200809 (10)B1908355.

Root Cause

Process change control

Action Taken

Greiner Bio-One North America Inc., notified customers on 03/12/2020 via "Urgent Product Recall" letter. The recall letter identified the affected products and requested the users to stop using the product and complete the product disposition site confirmation form. Consignees were asked to discard/destroy the affected products. If you have additional questions, please call our Technical Service at 800-815-8112.