RecallDepth

Centurion- Circumcision Tray Kits DYNDF1078C CIRCUMCISION TRAY DYNDF1082 CIRCUMCISION TRAY DYNDF1082 CIRCUMCISION TRAY DYNDF1093C UNIVERSAL CIRCUMCISION TRAY W/O CLAMP DYNDF1093C UNIVERSAL CIRCUMCISION TRAY W/O CLAMP DYNDF1109 TRAY,CIRCUMCISION DYNDF1128 CIRCUMCISION TRAY

This recall has been terminated (originally issued December 11, 2019).

Company
Centurion Medical Products Corporation
Recall Initiated
December 11, 2019
Terminated
September 4, 2020
Recall Number
Z-1051-2020
Firm Location
Williamston, MI
Official Source
View on FDA website ↗

Reason for Recall

Incomplete seals on the sterile package may compromise the sterility

Distribution

Nationwide Foreign: GHANA

Lot / Code Info

Lot Numbers:  DYNDF1078C CIRCUMCISION TRAY 2019051390 DYNDF1082 CIRCUMCISION TRAY 2019011590 DYNDF1082 CIRCUMCISION TRAY 2018121490 DYNDF1093C UNIVERSAL CIRCUMCISION TRAY W/O CLAMP 2019051490 DYNDF1093C UNIVERSAL CIRCUMCISION TRAY W/O CLAMP 2018111490 DYNDF1109 TRAY,CIRCUMCISION 2019020690 DYNDF1128 CIRCUMCISION TRAY 2018121490

Root Cause

Packaging process control

Action Taken

Centurion issued a notification emailed to Medline Industries, Centurion's parent company. Centurion issued mailing letters via First Class Mail beginning on 12/12/2019, letter states reason for recall, health risk and action to take: Please identify all inventory that you have within your possession and segregate it in a location where it cannot be accessed for use. 2) If you have transferred possession of this product to another individual, to another department or to another location in your organization, you are required by FDA regulations to notify them of this recall communication. Please ask that any affected product be returned to you for collection. 3) Affected product is not to be returned to Centurion, and should be destroyed/discarded. Document the destruction of all product on the enclosed accountability record. 4) Please complete the enclosed accountability record, whether or not you have product, and fax or email at your earliest convenience to the attention of Lisa Carpenter, Director of Quality Assurance Centurion Medical Products Corporation, (517) 546-5400 Ext.1122.

More recalls by Centurion Medical Products Corporation

View all recalls by this company →