Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide Product Code: ASK-04020-PMC The Arrow Arterial Catheterization device permits access to the peripheral arterial circulation or to other small vessels.

Company
Arrow International Inc
Recall Initiated
December 12, 2019
Recall Number
Z-0844-2020
Quantity
80 eaches
Firm Location
Reading, PA

Reason for Recall

The lidstock label incorrectly displays a MR (Magnetic Resonance ) Conditional symbol (instead of a MR Unsafe symbol) in reference to a component of the kit

Distribution

US distribution to SC

Lot / Code Info

Lot/Batch Number: PMC 23F19H0224

Root Cause

Labeling Change Control

Action Taken

Arrow International issued an " Urgent Medical Device Recall Notice" to its US consignee on 12th Dec 2019 via FedEx 2-day mail. The letter states reason for recall, health risk and action to take: If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. If you have no affected stock, please complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. Questions contact your local sales representative or Customer Service at 1-866-396-2111.

More recalls by Arrow International Inc

View all recalls by this company →