Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits: Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF 9735502-G02 (GTIN 00763000208462); Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169540521); and Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169641006) (Japan). The percutaneous pin adapter is a component of the StealthAir Frame Assembly.

This recall has been terminated (originally issued December 6, 2018).

Company
Medtronic Navigation, Inc.
Recall Initiated
December 6, 2018
Terminated
December 17, 2020
Recall Number
Z-0707-2020
Quantity
1,297 percutaneous pin adapters
Firm Location
Louisville, CO
Official Source
View on FDA website ↗

Reason for Recall

Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.

Distribution

Distribution was nationwide. There was government/military/foreign distribution.

Lot / Code Info

Percutaneous pin adapter lot numbers: 150114, 150505, 150506, 150904, 151005, 151007 160201, 160224, 160419, 160617, 160707, 170216, 180110, and 180409.

Root Cause

Device Design

Action Taken

The recalling firm issued a Dear Healthcare Professional letter dated 12/6/2018 via FedEx explaining the issue and requesting the consignee examine their inventory and quarantine the percutaneous pin adapter for return to the firm. Only the percutaneous pin adapter was to be returned and not the remaining items in the assembly.