RecallDepth

TriMed Peg Guide Extender, REF PEG-XTNDR, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.

Company
Trimed Inc.
Recall Initiated
September 28, 2018
Recall Number
Z-0358-2020
Quantity
1,903 Peg Guide Extenders
Firm Location
Santa Clarita, CA

Reason for Recall

The spring on the PEG-Extender can rust due to the type of steel used to manufacture it.

Distribution

Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.

Lot / Code Info

Lot number 2858, 2861, and 3749

Root Cause

Nonconforming Material/Component

Action Taken

The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.