TriMed Peg Guide Extender, REF PEG-XTNDR, nonsterile, Rx. The firm name on the label is TriMed Inc., Santa Clarita, CA.
- Company
- Trimed Inc.
- Recall Initiated
- September 28, 2018
- Recall Number
- Z-0358-2020
- Quantity
- 1,903 Peg Guide Extenders
- Firm Location
- Santa Clarita, CA
Reason for Recall
The spring on the PEG-Extender can rust due to the type of steel used to manufacture it.
Distribution
Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.
Lot / Code Info
Lot number 2858, 2861, and 3749
Root Cause
Nonconforming Material/Component
Action Taken
The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.