VITROS Immunodiagnostic Products - FSH Reagent Pack (10758750000302) 1931922
- Company
- Ortho Clinical Diagnostics Inc
- Recall Initiated
- July 12, 2018
- Recall Number
- Z-0059-2019
- Quantity
- 1955
- Firm Location
- Raritan, NJ
Reason for Recall
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
Distribution
Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.
Lot / Code Info
2470 2480 2490 2500 2510 2520 2530
Root Cause
Device Design
Action Taken
On 12 July 2018, a customer letter was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail to all customers who could process the affected products (i.e., VITROS 3600, 5600 and/or ECi/ECiQ Systems) to inform them of the potential interference of biotin at levels lower than stated in the IFUs and that Ortho is in the process of revising the IFUs for the affected products. Ortho recommended customers follow normal laboratory procedures for troubleshooting of samples containing other assay interferences and it was acceptable to continue using the affected products. If you have questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.