Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures

This recall has been terminated (originally issued July 13, 2018).

Company
Haemonetics Corporation
Recall Initiated
July 13, 2018
Terminated
June 16, 2020
Recall Number
Z-2950-2018
Quantity
432 units
Firm Location
Braintree, MA
Official Source
View on FDA website ↗

Reason for Recall

OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.

Distribution

US Nationwide Distribution

Lot / Code Info

Lot # 1017049 , 1150H-00, 1017072

Root Cause

Process control

Action Taken

The firm, Haemonetics, notified customers by an email titled "URGENT: MEDICAL DEVICE RECALL" on July 13, 2018 . The email described the product, problem, and actions to be taken. The customers were instructed to inspect product for defect and return units of the affected lots for replacement with non-affected units; complete the attached customer acknowledgement form via FAX TO +1-781-356-3558 OR SCAN AND E-MAIL TO: CORPORATEREGULATORY@HAEMONETICS.COM. You can contact your local customer service representative at: North America 1.800.537.2802 or email:customerservicena@haemonetics.com.