RecallDepth

NOMAD Pro2 Packaged X-Ray System, Part Number 08500021; NOMAD Pro2 Packaged X-Ray System Black, Part Number 08500022; NOMAD Pro2 Vet Packaged X-Ray System, Part Number 08500023; NOMAD Pro2 Vet Packaged X-Ray System Black, Part Number 08500024; NOMAD Pro2 Packaged X-Ray System, GBL, Part Number 08500025 and Part Number 08500026; NOMAD Pro2 Packaged X-Ray System, GLB, Vet, Part Number 08500027 and Part Number 08500028; NOMAD Pro2 Package X-Ray System, GLB, Black, Part Number 08500029

This recall has been terminated (originally issued September 2, 2016).

Company
Aribex Inc
Recall Initiated
July 18, 2016
Posted
September 2, 2016
Terminated
July 26, 2017
Recall Number
Z-2717-2016
Quantity
16,871
Firm Location
Charlotte, NC
Official Source
View on FDA website ↗

Reason for Recall

Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway condition as a result of a short circuit on the battery pack PCB.

Distribution

AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA. HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY , Puerto Rico, Afghanistan, Argentina, Australia, Bahamas, Bahrain, Belgium, Bermuda, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kuwait, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Saudi Arabia, Papua New Guinea, Philippines, Poland, Qatar, Saudi Arabia, Slovenia, South Africa, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, UAE, United Kingdom, Venezuela, Vietnam

Lot / Code Info

Part Number 08500021 - Serial Number Range XD-100921-05 - XD-150109-11, 10000 - 20811, XD-140929-39 - XD-150123-20; Part Number 08500022 - Serial Number Range XE-130528-03 - XE-150123-38, 10000 - 20541, XE-140908-32 - XE-150123-40; Part Number 08500023 - Serial Number Range XF-121008-05 - XF-150205-03, 10003 - 20032, XE-150205-01; Part Number 08500024 - Serial Number Range XG-130520-01 - XG-141210-05, 1000 - 20019; Part Number 08500025 - Serial Number Range XI-140303-02 - XI-150210-9, XI-150202-04 - XI-150210-19; Part Number 08500026 - Serial Number Range 10039 - 20041; Part Number 08500027 - Serial Number Range 1000 - 20006, Part Number 08500028 - Serial Number Range 10000 - 20030; Part Number 08500029 - Serial Number Range 10000.

Root Cause

Other

Action Taken

The firm sent a "FIELD ACTION NOTICE" to all affected customers.