Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434

This recall is currently active, issued June 17, 2021. It was issued by Medicalplastic S.R.L. Via Saverio Mercadante.

Company
Medicalplastic S.R.L. Via Saverio Mercadante
Recall Initiated
June 17, 2021
Recall Number
Z-0245-2022
Quantity
581 units
Official Source
View on FDA website ↗

Reason for Recall

Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.

Distribution

Memphis, TN

Lot / Code Info

Lot # 20LP02813, Serial #D147 to D149; Lot # 20LP02812, Serial # D115 to D119 ; Lot # 20LP02816, Serial # D099 to D114 ; Lot # 20HP01926, Serial # D059 to D065 ; Lot # 20LP02819, Serial # D175 to D198 ; Lot # 20LP02818, Serial # D129 to D146 ; Lot # 20LP02815, Serial # D120 to D128 ; Lot # 20MP03078, Serial # D374 ; Lot # 20HP01927, Serial # D278 to D282 ; Lot # 20MP03037, Serial # D294 to D296 ; Lot # 20LP02827, Serial # D199 to D258 ; Lot # 20 MP03036, Serial # D288 to D293 ; Lot # 20MP03077, Serial # D370 to D373 ; Lot # 20MP03035, Serial # D259 to D277 ; Lot # 20MP03041, Serial # D283 to D287 ; Lot # 20MP03049, Serial # D343 to D369 ; Lot # 20MP03048, Serial # D297 to D342 ; Lot # 20MP03123, Serial # D375 to D459 ; Lot # 20HP01928, Serial # D460 to D474 ; Lot # 20MP03126, Serial # D482 to D491 ; Lot # 20MP03125, Serial # D492 to D522 ; Lot # 20MP03124, Serial # D475 to D481 ; Lot # 21BP00265, Serial # D523 to D574 ; Lot # 21BP00266, Serial # D626 to D672 ; Lot # 21BP00271, Serial # D588 to D625 ; Lot # 21BP00272, Serial # D575 to D587 ;

Root Cause

Process control

Action Taken

The firm notified its sole consignee of the issue and its intent to recall by email on 06/17/2021. The recalled washers will be replaced.