LeadCare II Blood Lead Test Kit, containing LeadCare II Lead Controls Product Usage: IVD test kit for the detection of lead in blood. The kit consists of sensors, treatment reagent, lead controls (two levels), heparinized capillary tubes and plungers, transfer droppers and a calibration button. The LeadCare II Lead Controls packaged in the kit lots listed in #2 below are the subject of this recall (Field Correction) action.

This recall has been terminated (originally issued August 15, 2017).

Company
Magellan Diagnostics, Inc.
Recall Initiated
January 20, 2016
Posted
August 15, 2017
Terminated
February 21, 2018
Recall Number
Z-3035-2017
Quantity
12,931 kits
Firm Location
North Billerica, MA
Official Source
View on FDA website ↗

Reason for Recall

LeadCare II Test Kits controls out of range

Distribution

Worldwide - US Nationwide distribution and the countries of: COLOMBIA, CANADA, THAILAND, INDIA, UAE, AUSTRALIA, GERMANY, CANADA, SOUTH AFRICA, SPAIN, NEW ZEALAND, KOREA

Lot / Code Info

Lot # :1507N Exp Date: 1/27/17; Lot # :1508N Exp Date: 2/27/17; Lot # :1510N Exp Date: 4/19/17; Lot # :1511M Exp Date: 5/3/17 Lot # :1511N Exp Date: 5/17/17 Lot # :1512M Exp Date: 6/2/17

Root Cause

Device Design

Action Taken

Magellan notified customers initially by Email, fax and/or verbal communication with Magellan Product Support representatives beginning on January 20, 2016 to advise of the problem and provide revised Assignment Control Value Matrix..