Endoskeleton TC Interbody System with Titan nanoLOCK Surface Technology, 6 Degree, Large, 8 mm, Product Number 5146-1608-N - Product Usage: indicated to be used with supplemental fixation cleared by the FDA for use in the cervical spine and autograft bone, allograft bone comprised of cancellous and/or corticocancellous bone, demineralized allograft with bone marrow aspirate, or a combination thereof.
- Company
- Medtronic Sofamor Danek Usa Inc
- Recall Initiated
- March 17, 2021
- Terminated
- May 3, 2023
- Recall Number
- Z-1440-2021
- Quantity
- 11 devices
- Firm Location
- Memphis, TN
Reason for Recall
The shelf box identifies a small device actually contains a large device, and the shelf box identifying a large device actually contains a small device.
Distribution
US Nationwide distribution.
Lot / Code Info
GTIN 00191375013167, Lot Number TM0126572
Root Cause
Packaging process control
Action Taken
The firm initiated the recall by letter on 03/19/2021, The letter requested that the consignee locate and isolate the impacted product and contact the Medtronic Sales Representative to coordinate the return and replacement of the device.
More recalls by Medtronic Sofamor Danek Usa Inc
Catalyft PL & PL40 Expandable Interbody System labeled as follows with corresponding model numbers:...
Nov 21, 2025
Grafton DBM Orthoblend, Model Numbers: a) T44125INT, b) T44135, c) T44145, d) T44150; Demineralized...