BD Hypodermic Syringe 10 ml, eccentric tip. Catalog (Ref No.) 305482

Company
Becton Dickinson & Company
Recall Initiated
December 23, 2020
Terminated
November 21, 2022
Recall Number
Z-1109-2021
Quantity
155,200 needles
Firm Location
Franklin Lakes, NJ

Reason for Recall

The product packaging for the 10 ml eccentric tip syringe may exhibit open seals. An open seal impacts package integrity and potentially compromises sterility of the syringe.

Distribution

Distribution US nationwide and Canada

Lot / Code Info

Catalog (Ref No.) 305482 100/package  UDI 00382903054824  Lot # 8142303  exp date 31-May-2023

Root Cause

Process control

Action Taken

On 01/13/2020 BD issued an Urgent Medical Device Recall notices to distributors via letters and via e-mail. Distributors are requested to identify their customers and provide BD a customer list so that BD can undertake further communication to the users. BD will distribute the recall communication to the distributors customers via FedEx upon receiving the customer information from the distributors. A copy of the BD product recall customer letter will be provided to distributors, in the case they opt to contact their customers directly.