RecallDepth

Flow-i Anesthesia System Product Usage: The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.

This recall has been terminated (originally issued February 22, 2016).

Company
Maquet Cardiovascular Us Sales, Llc
Recall Initiated
December 10, 2015
Posted
February 22, 2016
Terminated
July 6, 2016
Recall Number
Z-0804-2016
Quantity
7 units within US
Firm Location
Wayne, NJ
Official Source
View on FDA website ↗

Reason for Recall

Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues. Only the Model C20 is affected in the US.

Distribution

US Nationwide in the states of CA, NY, and OR

Lot / Code Info

Model numbers: C20, C30 and C40 (Only Model # C20 for US)  Part numbers: 66 77 200 (US Only), 66 77 300 and 66 77 400  Serial numbers (2743, 2852, 2854, 2855, 2856, 2858

Root Cause

Nonconforming Material/Component

Action Taken

Maquet sent "Urgent Medical Device Field Action Field Correction For Flow-i Anesthesia Systems" letters/Field Correction Response forms dated 12/10/2015 via UPS to affected customers. The notification informed customers of the issue, asked them to distribute the notification to appropriate staff; verify the serial number(s) and informed them they could continue to use the system if they were not experiencing electrical issues. Customers were informed that a Maquet service representative would be contacting them to arrange a time to replace their affected units. Customers were asked to complete and return the enclosed Field Correction Response Form to Maquet at FieldActions@maquet.com or by fax to 973.860.1675. If the customer had any related questions they were instructed to contact Maquet Technical Support at 1.888.627.8383 (option 3; option 1; option 1) Monday through Friday between 8:00 am and 5:00 pm EST.

More recalls by Maquet Cardiovascular Us Sales, Llc

View all recalls by this company →