Olympus, Model No. CF-Q140L, flexible scope, large diameter, video colonoscope

This recall has been terminated (originally issued January 20, 2016).

Company
Integrated Medical Systems Inc
Recall Initiated
January 20, 2016
Terminated
October 13, 2016
Recall Number
Z-0909-2016
Quantity
2 instruments
Firm Location
Birmingham, AL
Official Source
View on FDA website ↗

Reason for Recall

If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.

Distribution

Distributed US (nationwide) and in Australia.

Lot / Code Info

Serial Numbers: 2913081,2600279

Root Cause

Nonconforming Material/Component

Action Taken

The firm notified their consignees of the issue by Fed Ex on 01/20/2016. The firm plans to coordinate with the medical facilities for inspection of the endoscopes.