RecallDepth

3.5mm Non Locking Cortical Screw, in the PediLoc Tibial Plate System Part number 00-0903-2510, 3.5mm x 10mm Non Locking Cortical Screw. Used in pediatric orthopedics.

This recall has been terminated (originally issued September 18, 2015).

Company
Orthopediatrics Corp
Recall Initiated
July 28, 2015
Posted
September 18, 2015
Terminated
March 17, 2016
Recall Number
Z-2796-2015
Quantity
5
Firm Location
Warsaw, IN
Official Source
View on FDA website ↗

Reason for Recall

The 00-0903-2510, 3.5mm x 10mm Non Locking Cortical Screw, Lot # 150192-J package may contain a different screw. The screw in the package is a 00-0903-2526, 3.5mm x 26mm, and is etched correctly.

Distribution

Domestic: TN, IA Foreign: England VA/DOD: None

Lot / Code Info

Part number 00-0903-2510, Lot number 150192-J

Root Cause

Packaging process control

Action Taken

OrthoPediatrics has initiated a Voluntary RECALL of Part # 00-0903-2510, 3.5mm Non Locking Cortical Screw, Lot # 150192-J, initially by contacting customers via telephone, and then by e-mailing notification letters on 7/28/2015. Customers returned the 5 mislabled screws, which were repacked with correct labeling. All recall activities were complete as of 8/14/2015.