Lithium Polymer Fiber Optic Module Batteries used in the Intera MR System upgraded to CDAS, Part Number: 9065 and 9067. These rechargeable batteries are used to supply power to the following products: Wireless ECG Module used with the Precess MRI compatible Patient Monitoring System, Wireless SpO2 Module used with the Precess MRI Compatible Patient Monitoring System, Philips Basic Triggering Set, Philips Monitoring Sensor Set, Philips Battery Set

This recall has been terminated (originally issued March 20, 2015).

Company
Philips Medical Systems, Inc.
Recall Initiated
November 20, 2014
Posted
March 20, 2015
Terminated
January 23, 2017
Recall Number
Z-1307-2015
Quantity
79
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

VCG battery ignited in a VCG unit when recharging.

Distribution

Distributed in the following states: AL, CA, FL, GA, IL, MA, MN, NC, ND, NE, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI; OUS: Australia, Austria, Belgium, China, Ecuador, Finland, France, Germany, Italy, Japan, Mexico, Netherlands, slovenia, South Africa, Sweden, Switzerland, United Kingdom

Lot / Code Info

Batteries with date codes 2808 or earlier.

Root Cause

Component design/selection

Action Taken

Consignees were informed of the recall via letter sent on November 20, 2014 via certified mailing. Field Service Engineers (FSE) will replace all affected batteries.