Synthes 130 Degree Aiming Arm and 125 Degree Aiming Arm; The 130 Degree Aiming Arm and 125 Degree Aiming Arm are a part of the TFN-ADVANCED (TFNA) System. The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures in adults and adolescents (12-21) in which the growth plates have fused.

This recall has been terminated (originally issued March 30, 2015).

Company
Synthes, Inc.
Recall Initiated
February 24, 2015
Posted
March 30, 2015
Terminated
September 21, 2015
Recall Number
Z-1352-2015
Quantity
131
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

The guide sleeve has the potential to jam in the Aiming Arms, which may delay disassembly and removal of instruments.

Distribution

US Nationwide Distribution - NY, TX, MD, OR, MS, ND, CA, AZ, MI, IL, PA, TN, AR, SD, SC, FL, VA, OK, and GA

Lot / Code Info

Part Numbers: 03.037.013, 03.037.014, Lot Numbers: 8903848, 9124017, 903851, 9124019, 8918566, 9124022, 9124012, 9124023, 9124014, 9124026,  9124016, 9124027, 8918563

Root Cause

Component design/selection

Action Taken

DePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated March 2, 2015, to affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to review inventory for the affected parts and following the actions provided in the letter for removal and replacement of these devices. For questions call 610-719-5450.