Edwards Lifesciences Swan-Ganz Thermodilution Catheter REF: 139F75 For use inpatients who require hemodynamic monitoring. Intended to be used in combination with clinical pressure monitoring equipment to measure right heart and pulmonary artery pressures, and with the Vigilance Monitor to measure cardiac output.

This recall has been terminated (originally issued April 17, 2009).

Company
Edwards Lifesciences, Llc
Recall Initiated
February 9, 2009
Posted
April 17, 2009
Terminated
April 11, 2011
Recall Number
Z-1238-2009
Quantity
35 units
Firm Location
Irvine, CA
Official Source
View on FDA website ↗

Reason for Recall

Cardiac output measurements were unavailable and the following message is displayed in the Vigilance monitor: "Fault. CCO Catheter Verification. Use Bolus Mode".

Distribution

Worldwide Distribution -- AL, AR, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Hawaii, and Puerto Rico and countries of Canada, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, UK, Greece, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Portugal, Sweden, Colombia, Australia, and Singapore.

Lot / Code Info

Lot: 58619364 - expiration date between May 2010 and November 2010.

Root Cause

Software design

Action Taken

Edwards initiated the worldwide recall via Urgent Customer Notification letters dated February 9, 2009 and are being sent by Federal Express starting on February 10, 2009. These letters are requesting that customers cease using Swan-Ganz" Continuous Cardiac Output (CCO) Thermodilution Catheter products from the indicated lots and that they return all unused product. Edwards Lifesciences requested that the customer provide written verification of stock and identify any unused product to be returned. Questions should be directed to Technical Service at 800-822-9837.