Steris C1160 Universal Flexible Processing Tray, for use with STERIS SYSTEM 1 Processor, used for sterilization of flexible endoscopes.

Company
Steris Corp
Recall Initiated
February 19, 2008
Posted
August 5, 2008
Terminated
February 24, 2012
Recall Number
Z-1190-2008
Quantity
28,766 units
Firm Location
Mentor, OH

Reason for Recall

Failure to effectively sterilize: Under certain conditions, the design/operation of the C1160 Tray may compromise the ability of the SYSTEM 1 processor to correctly assess the state of the high pressure pump during the Sterile Cycle. The L6 sensor fails to alarm when the high pressure pump is not working properly.

Distribution

Worldwide Distribution, including USA, Puerto Rico, Canada, and Japan.

Lot / Code Info

All codes

Root Cause

Device Design

Action Taken

Steris sent an Urgent Safety Alert/Recall letter dated 2/8/2008, to all customers. The letter stated that any device sterilized in the C1160 Tray should not be used clinically prior to a completion of a successful diagnostic cycle. Until further notice, you must run a diagnostic cycle after each sterile cycle involving a C1160 tray or alternately you must discontinue use of these trays. Steris service technicians will schedule visits to correct the problem.