Osteo-Clage Cable System Package Assembly (01-0001-S). Product labeled in part, "2.0mm x 850 mm CoCr Cable / Sleeve... REF 01-0001-S LOT W59192... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR 97124-9432 ".

This recall has been terminated (originally issued August 22, 2008).

Company
Acumed Llc
Recall Initiated
March 28, 2008
Posted
August 22, 2008
Terminated
February 3, 2010
Recall Number
Z-2140-2008
Quantity
50 units
Firm Location
Hillsboro, OR
Official Source
View on FDA website ↗

Reason for Recall

Cables of incorrect diameter were used in product assembly, resulting in the sleeve not being able to bind to the cable after crimping.

Distribution

Nationwide Distribution --- including states of AR, CA, PA, GA, KS, MO, and TN.

Lot / Code Info

Work order number/Lot: W59192 (W=work order, ###=sequential number); Expiration date: 2015-01 (Year/Month)

Root Cause

Employee error

Action Taken

On 3/28/08, the firm sent an Urgent Notification letter to their consignees. The letter explained the issue and requested the customers return the product to Acumend. On 7/16/08, the firm sent a 2nd notification, via mail, to thier customers. The Urgent Notice: Device Recall 2nd Notification, advised customers of the recall, the hazard and requested customers to return existing product to Acumed.