Siemens Medical Solutions USA, Inc., Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System.

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
October 14, 2011
Posted
November 23, 2011
Terminated
October 16, 2012
Recall Number
Z-0284-2012
Quantity
3 units
Firm Location
Hoffman Estates, IL

Reason for Recall

Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo MI Applications version VA60A with Foresight detectors firmware version v1.5.9.12 on the Symbia systems.

Distribution

Florida, Illinois and Texas

Lot / Code Info

Symbia T Series systems, serial numbers 12316, 1236 and 1013, both with syngo MI Applications version VA60A and Foresight Detector AEB firmware version v1.5.9.12

Root Cause

Component change control

Action Taken

Siemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated October 14, 2011 to all affected customers. The letters informed customers of the acquisition software defect in the energy window position when an isotope is peaked, which could potentially impact image interpretation, and that the AEB firmware version on their imaging system has been changed to a previously released version until acquisition software service pack is available. For questions or assistance they were instructed to contact Siemens Medical Solutions USA, Inc. at (800) 888-7436.