Oxylog 3000 Emergency and Transport Ventilator Continuous Respirator with software version 1.10 pre-installed; Catalog number 2M86300.

This recall has been terminated (originally issued October 27, 2008).

Company
Draeger Medical, Inc.
Recall Initiated
October 3, 2008
Posted
October 27, 2008
Terminated
October 16, 2009
Recall Number
Z-0202-2009
Quantity
50 units
Firm Location
Telford, PA
Official Source
View on FDA website ↗

Reason for Recall

May experience an interruption of ventilation for approximately 5 seconds

Distribution

Nationwide Distribution --- including states of AZ, CA, FL, IL, IN, KY, LA, ME, MI, MO, NC, NE, NJ, NY, OH, PA, TX, UT, VT, WA, WV, and WY.

Lot / Code Info

Serial numbers SRZA-0182 and higher tahn 0182, SRZB-xxxx, SRZC-xxxx, SRZD-xxxx, SRZE-xxxx, SRZF-xxxx, SRZH-xxxx, and SRZJ-xxxx. Note that xxxx stands for numbers 0001 through 9999.

Root Cause

Software design

Action Taken

The firm issued an Urgent - Medical Device Recall letter dated September 2008. Software version 1.10 may under specific conditions experience an interruption of ventilation for approximately 5 seconds. During that time period, the patient airway pressure may drop to zero. This phenomenon was identified by Draeger engineers and to date has not been reported by any user. The firm's investigations revealed that when ventilating in BIPAP mode with the pressure rise time (Ramp) set to the highest setting and the expiration time set to exactly 0.5 seconds, the inspiration phase can coincide with the internal auto zero function of the device. Under these circumstances a pressure spike will occur upon reaching Pmax +5mbar (cmH2O) which with software version 1.10 will lead to the inspiratory phase stopping. Ventilation will resume after approximately 5 seconds. The device will generate a Paw Measurement INOP alarm. Until the new software is available, customers are being instructed to avoid the use of the following combination of settings: 1) BIPAP mode; 2) Ramp = highest, and 3) Texp = 0.5 seconds.