RecallDepth

Cotton-Leung Biliary Stent., CLSO-10-5, Stent Diameter: 10 FR, Stent Length: 5cm, Disposable-Single use only, Rx Only, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105. Used to train obstructed biliary ducts.

This recall has been terminated (originally issued September 23, 2008).

Company
Cook Endoscopy
Recall Initiated
July 2, 2008
Posted
September 23, 2008
Terminated
June 4, 2009
Recall Number
Z-2331-2008
Quantity
20 units
Firm Location
Winston Salem, NC
Official Source
View on FDA website ↗

Reason for Recall

The product packaging label indicates the stent is 5cm, but the actual stent inside the packaging measures 10cm.

Distribution

Worldwide Distribution --- USA including states of MO, NC, NV, TX, and UT and country of India.

Lot / Code Info

Lot number W512494

Root Cause

Process control

Action Taken

Consignees were notified on/about 07/11/2008 via an Urgent - Product Recall Notice. All medical user consignees were instructed to review the attached list and quarantine any affected product in their possession. They are to return any affected product to Cook Endoscopy. A reply form was attached to verify receipt of notification and quantity product returned. All product returned as part of this recall initiative will be dispositioned to scrap and destroyed.

More recalls by Cook Endoscopy

View all recalls by this company →