Hologic Oasis Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is indicated for the estimate of bone mineral density, comparison of measured variables obtained from a given scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

This recall has been terminated (originally issued September 17, 2008).

Company
Hologic, Inc.
Recall Initiated
June 27, 2008
Posted
September 17, 2008
Terminated
October 25, 2010
Recall Number
Z-2188-1991
Quantity
15 units
Firm Location
Bedford, MA
Official Source
View on FDA website ↗

Reason for Recall

Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate.

Distribution

Nationwide Foreign: Canada Algeria, AE, Australia, Belgium,China,France, Germany, Hungary, Hong Kong, Iceland, Iran, Ireland, Israel, Italy,Japan,Korea, Kuwait, Lithuania, Mexico, Myanmar, Netherlands, Pakistan, Peru, Russia, S. Africa, Saudi Arabia, Singapore, Slovak Rep., Switzerland, Syria, Taiwan, UK

Lot / Code Info

S/N: 11/01/06 B 1/04/08B 11/01/06 A 6/24/05-A 9/25/06-01 9/25/06-02 9/25/06-03 9/25/06-04 9/28/06-01 9/28/06-02 9/28/06-03 9/12/07B 10/02/07B 10/02/07A and 1/04/08A

Root Cause

Software design

Action Taken

Hologic notified customers by registered letter titled Urgent: Medical Device Recall, dated 6/27/08, providing instructions to update the software to Version 2.3 and discard version APEX 2.0, 2.1 and 2.2 manuals. If you have questions or require assistance with the software installation, contact Hologic's Help Desk at 1-800-321-4659.